Total cards: 24•Content version: v4•Public deck•Updated: 8/1/2026

Card preview (24)

#1
Front (Question)

What method was used in the lab to measure total protein?

Back (Answer)

Biuret method

#2
Front (Question)

What does the biuret method consist of?

Back (Answer)

Copper sulfate reacting with peptide bonds with increasing intensity to their concentration. Colored chelate forms with amine nitrogen of the peptide bond with an absorbance at 540nm.

#3
Front (Question)

What type of amino acids react with the biuret method?

Back (Answer)

Tripeptides and polypeptides. Dipeptides or single amino acids do not.

#4
Front (Question)

What is intensity of color in biuret method results proportional to?

Back (Answer)

Number of peptide bonds and therefore protein concentration

#5
Front (Question)

Identify the other methods for assessing total protein.

Back (Answer)

Kjehldahl Refractometry UV absorption

#6
Front (Question)

What method is used for detecting albumin?

Back (Answer)

Dye binding technique

#7
Front (Question)

How does the dye binding technique work?

Back (Answer)

In dye binding technique, anionic dye binds to albumin at a pH of 4.2, pH at which albumin is cationic. BCG binds specifically to albumin although it will bind other plasma proteins over time. BCG binds to protonated amine groups on the albumin molecule and the complex forms a color at 550-630 nm

#8
Front (Question)

What type of specimens are used for total protein albumin measurements?

Back (Answer)

Serum Heparinized plasma Body fluids/urine

#9
Front (Question)

What is the reference interval for total protein?

Back (Answer)

6.5-8.3 g/dL

#10
Front (Question)

What is the reference interval for albumin?

Back (Answer)

3.5-5.5 g/dL

#11
Front (Question)

What is automation?

Back (Answer)

The mechanization of chemical analyses done in order to minimize manual manipulation

#12
Front (Question)

List the steps in the automated process

Back (Answer)

Separation/mixing Sample/reagent handling Incubation Reaction and reaction analysis Derivation/interpretation of results Result reporting

#13
Front (Question)

What are the advantages of automation?

Back (Answer)

Time saving Removes human performance variability Reduces error Saves money

#14
Front (Question)

What are the disadvantages of automation?

Back (Answer)

High initial/startup cost Discontinued or updated products and reagents Special training required

#15
Front (Question)

What are discrete analyzers?

Back (Answer)

Analyzers in which samples and reagents are mixed and react in individual cucette or slide

#16
Front (Question)

What are continuous flow analyzers?

Back (Answer)

Analyzers in which sample is injected into a reagent stream and reactions take place in tubing

#17
Front (Question)

What can be a major problem with continuous flow analyzers?

Back (Answer)

Carryover

#18
Front (Question)

How does sample processing with batch analyzers work?

Back (Answer)

Only one analysis is performed on a single specimen at a time or many specimens are sequentially moved through the analyzer as one batch

#19
Front (Question)

How does random access analyzers work?

Back (Answer)

Operators can program a number of tests, profiles/panels, QC on any given specimen in any order. Many analyses can be performed on a single specimen before moving on to the next sample

#20
Front (Question)

What is the most common system of analyzers?

Back (Answer)

Random access analyzers

#21
Front (Question)

What are calibrators or standards?

Back (Answer)

These are substances of known concentration run initially to create a standard curve

#22
Front (Question)

How many cakibartions are there?

Back (Answer)

One calibration per analyte

#23
Front (Question)

What are unknowns comapred to?

Back (Answer)

Standards on the curve once validated through calibration and QC

#24
Front (Question)

What are controls?

Back (Answer)

Substances of a known general range run daily or per shift to confirm system validity through QC