#1
正面 (问题)

Used to process priority stat samples and the fastest way to process a sample.

背面 (解答)

STAT table

#2
正面 (问题)

the sample probe/liquid level detector dispenses samples and or diluent, and detects liquid level. Sample is aspirated from the tube/cup and dispensed into the cuvette wheel

背面 (解答)

Sample transfer unit

#3
正面 (问题)

Two reagents probe/liquid level detectors dispense reagent and or diluent, and detect liquid level. The probe assemblies are designed for up and down collision detection. Reagent is aspirated from R1 and R2 compartment and dispensed into the cuvette wheel.

背面 (解答)

Reagent transfer unit

#4
正面 (问题)

Mix bars used to mix the reagent and sample in the cuvette

背面 (解答)

Mixing unit

#5
正面 (问题)

Consist of a halogen lamp, lenses, a diffraction grafting, and a photodetector to measure the amount of light transmitted through the reaction solution in the cuvette

背面 (解答)

Photometry unit

#6
正面 (问题)

Cleans, rinses, and dries the cuvettes automatically

背面 (解答)

wash nozzle unit

#7
正面 (问题)

Define ISE unit

背面 (解答)

Diluted samples is passed through sodium, potassium, and chloride ion selective electrode to determine the concentration by comparing the electrical potential different to the reference electrode.

#8
正面 (问题)

The screen to check system messages relating to sample status and analyzer status.

背面 (解答)

Home outline

#9
正面 (问题)

Check analyzer status menu: Blue

背面 (解答)

no error

#10
正面 (问题)

Check analyzer status menu: yellow or orange

背面 (解答)

non fatal error. the analyzer can be started

#11
正面 (问题)

Check analyzer status menu: Red

背面 (解答)

Fatal error. The analyzer can’t be started

#12
正面 (问题)

3 important sample requirements for the instrument to sample correctly.

背面 (解答)

• should not contain fibrous material or fibrin, except for WB. • No air bubbles • Dispense sample volume in the quantity required for analysis plus the dead volume

#13
正面 (问题)

Rack colors: white

背面 (解答)

used for analyzing routine samples

#14
正面 (问题)

Rack colors: Blue

背面 (解答)

used for calculation of reagent blanks for creation of calibration curves

#15
正面 (问题)

Rack colors- yellow

背面 (解答)

used for creating of calibration curves

#16
正面 (问题)

Rack colors- green

背面 (解答)

used for analysis of QC samples

#17
正面 (问题)

Rack colors- orange

背面 (解答)

used for manual repeat run analysis

#18
正面 (问题)

Rack colors- red

背面 (解答)

used for emergency sample analysis

#19
正面 (问题)

Daily variation chart for QC- green

背面 (解答)

normal data

#20
正面 (问题)

Daily variation chart for QC- yellow

背面 (解答)

data outside the range of the single check level programmed in QC specific

#21
正面 (问题)

Daily variation chart for QC- orange

背面 (解答)

abnormal data (data not included in the QC statistical values)

#22
正面 (问题)

What colors do the operator observe on the analyzer maintenance screen if the analyzer requires maintenance and what are the meanings of the colors?

背面 (解答)

Orange- over due Yellow- expiring The colors for daily maintenance appear 3 hours before the maintenance check.

#23
正面 (问题)

List 3 reasons that teh instrument may not be able to calculate a result

背面 (解答)

• mechanical error has occurred • outside fo cuvette walls or the cuvette wheel is wet • in the rate rxn, fewer than 3 photometric readings statisfy the assay creteria specified in teh specific test parameters

#24
正面 (问题)

List the first possible causes and corrective actiosn to be taken for the errors

背面 (解答)

sample can be severly lipemic, icteric, hemolytic or can contain excessively large amount of the analyte being tests. dilute the sampel and run the test again.