MIDTERM LEC 1.q
No description yet. This curated deck is ready for daily study.
Card preview (24)
Checks if the instrument measures light accurately (wavelength and absorbance)
Quality Assurance for Spectrophotometers
Makes sure the readings are linear and detects stray light
Quality Assurance for Spectrophotometers
Wavelength accuracy means the correct light color passes through the device
Quality Assurance for Spectrophotometers
Quality Assurance for Spectrophotometers Uses special filters to check: (2)
Didymium filter (peak at 600 nm) Holmium oxide filter (peak at 360 nm)
Neutral density filters and dichromate solution test if absorbance is linear
Quality Assurance for Spectrophotometers
Uses filters or solutions with known absorbance at specific wavelengths
Absorbance Check
Operator measures and compares results to known values
Absorbance Check
Best for absorbance check
below 2.0
Absorbance Check Absorbance _ _ cause inaccurate results and bending of the linear plot (deviation from Beer’s law)
above 2.0
It states that the concentration of the unknown substance is directly proportional to the absorbed light (absorbance or optical density) and inversely proportional to the amount of transmitted light (% Transmittance).
Beer’s Law
It mathematically establishes the relationship between concentration and absorbance.
Beer’s Law
It is the amount of light absorbed
Absorbance (A)
It is proportional to the inverse log of transmittance.
Absorbance (A)
It is mathematically derived from % T.
Absorbance (A)
Absorbance (A) formula
A = abc = 2 - log%T
Absorbance (A) where: A = _ a = _ absorptivity; absorptivity of the compound under standard conditions b = _ of light through the solution c = _ of absorbing molecules/solution
Absorbance Molar Length Concentration
It is the ratio of the radiant energy transmitted (T) divided by the radiant energy incident on the sample (I).
Percent Transmittance (%)
The %T measured by commercial spectrophotometers is the ratio of the sample transmitted beam divided by the blank transmitted beam (sample beam signal/blank beam signal × 100).
Percent Transmittance (%)
In actual practice, the light transmitted by a blank is substituted for _
IO= Intensity of light striking the sample
Blanking Technique Contains reagents only; corrects for reagent color absorbance
Reagent Blank
Blanking Technique Contains sample without reaction product; corrects for sample interference (ex. hemoglobin)
Sample Blank
Blanking Technique Contains serum only; used to measure background absorbance.
Blank without reagent
process may not be effective in some cases of turbidity, and ultracentrifugation may be necessary to clear the serum or plasma of chylomicrons.
Blanking Technique
Blanking Technique interfere mainly by increasing light scatter (turbidity).
Lipids